37
4
Operation
4.7 Trendelenburg
WARNING
When you operate the Trendelenburg operation with a patient on the operating table,
The patient may fall from the operating table.
CAUTION
When performing the head down operation, be careful
may get damaged.
When performing the head down operation, be careful
base. It may get damaged.
Head plate
fixing knob
Reverse Trendelenburg (Head up)
1.
Press
.
The tabletop moves to the head up position.
Head up
Trendelenburg (Head down)
1.
Press
.
The tabletop moves to the head down position.
Head down
NOTE
The maximum angle of
head up is 30 degrees from
the level position.
30°
30°
When Trendelenburg is
operated in the opposite
direction from the current
Trendelenburg position, it
will stop in the level position.
Once it stops, release the
switch and press it once
again; the tabletop will be
tilted to the maximum angle.
NOTE
The maximum angle of head
down is 40 degrees from the
level position.
40°
40°