Before contacting for repair
s
8
. Before contacting for repairs
Checking causes and countermeasures
The following problems can occur even if the operating table is not malfunctioning. Check the following points
before requesting repairs.
Status
Possible cause
Measures
The table cannot be
turned on.
The connector of the control unit is not
connected properly
Insert the connector completely.
( Page 20)
The battery is fully discharged.
Charge the battery. ( Page 33)
A function button on
the control unit does
not function.
The connector of the control unit is not
connected properly.
Insert the connector completely.
( Page 20)
You did not press the E switch before the
function switch.
switch. ( Page 36 to 52)
Protection circuit is activated.
Wait for about 30 minutes to operate.
( Page 70)
The brake button on
the control unit does
not function.
The emergency brake release handle is
in a release (UNLOCK) position.
Turn the emergency brake release
handle toward “LOCK”. ( Page 69)
After releasing the brake by the emergency
brake release handle, the brake has not
been unlocked by the control uni t.
Unlock the brake by the control unit.
( Page 69)
The back plate cannot
be lowered below the
level position.
The tabletop is slid in the foot direction
beyond the standard center position.
Slide the center of the tabletop in the
head direction beyond the standard
center position. ( Page 41)
The tabletop cannot be
slid in the foot direction.
Back plate is lowered below the level
position.
Move up the back plate from the level
position. ( Page 39)
cannot be operated.
The tabletop is slid in the foot direction
beyond the standard center position.
Slide the center of the tabletop in the
head direction beyond the standard
center position. ( Page 41)
The leg plate is not
lowered from the level
position.
The tabletop is in the possible
interference area of the leg plate.
Slide the tabletop toward the leg side.
( Page 41)
If the situation does not improve even if the above countermeasures are implemented, request repairs from
your distributor or Mizuho
In case of malfunction
Implement the follow measures when the operating table is broken.
1.
Turn the power switch off and disconnect the power cord from the medical grade outlet.
2.
Place an "Out of Order" or "Do Not Use" sign on the operating table.
WARNING
providers. Make sure to contact your distributor or Mizuho for maintenance or repairs.
Do not disassemble the operating table. Unauthorized disassembling may cause a
In order to prevent infections, make sure to clean and disinfect the operating table
when requesting to have it repaired.