Coniguration PR 5900
[Operating] - [System setup] - [Weighing points] - [Weighing point X] - [xBPI scale] - [In-
terface] - [Built-in RS
‑
232]
Note:
For further information, see the platform scale operating instructions.
4.5.8.4
Connecting a Mettler scale via RS
‑‑
232
A Mettler scale with MT-SICS protocol can be connected via the internal RS
‑
232 interface.
Example: Type XS6002SDR
PR 5900
PR 5900
RS-232
TxD
RxD
CTS
GND
RTS
5
2
3
7
8
TxD
RxD
GND
CTS
RTS
1
①
Mettler scale type XS6002SDR
MT-SCIS protocol (slave)
Coniguration PR 5900
[Operating] - [System setup] - [Weighing points] - [Weighing point X] - [Mettler-Scale] -
[Interface] - [Built-in RS
‑
232]
For the parameter settings for the Mettler scale, please refer to [System setup] -
[Weighing points] - [Mettler-Scale] - [Parameters] in the PR 5900 operating instructions.
4 Device installation
Process Controller Maxxis 5 PR 5900
Minebea Intec
EN-51