Note:
The EtherNet/IP card PR 1721/77 is supported by software version 2.24 and device serial
number 30363xxxxx or higher.
The IP address and subnet mask are set under [Fieldbus parameters] (see the
PR 5900 operating manual under [System setup] - [Fieldbus parameters] -
[EtherNet IP] )
The EDS ile is stored on the CD supplied with the device (ieldbus directory of the
respective device). The current ile is also available to download online:
http://www.minebea-intec.com
4.6.12.1
LEDs
PR 1721/57
Identii-
cation
Description
NS
Network status LED
MS
Module status LED
Link
Link/Activity LED
Network status (NS)
LED status
Description
Comments
Of
Module is oline
-
No power
-
No IP address
Constant green
Module is online
One or more connections exist (CIP
class 1 or 3).
1 Hz lashing green
Module is online
No connection
Constant red
Critical connection error Duplicate IP address
1 Hz lashing red
One or more connections has excee-
ded the connection time (CIP class 1
or 3).
Module status (MS)
LED status
Description
Comments
Of
Module is uninitialized
-
No power
-
Module has the status "SETUP" or
"NW_INIT".
Constant green
Module is initialized
Normal operation
1 Hz lashing green
Error after test
Error occurred after test.
2 Hz lashing green
This is used for node identiication in
the network.
Process Controller Maxxis 5 PR 5900
4 Device installation
EN-118
Minebea Intec