4.5.8
RS-232 interface (internal)
The device is equipped with an integrated RS-232 interface.
This interface is conigurable, and can be used, for example, for data transmission to a
remote display or printer.
Connection options using the RS-232 interface:
-
Connecting a YDP14IS ticket printer via RS-232; see Chapter
-
Connecting additional printers; see Chapter
-
Connecting an IS platform via RS-232; see Chapter
-
Connecting a Mettler scale via RS-232; see Chapter
GND CTS RxD TxD RTS
Technical data
Description
Data
Connection
Terminal, 5-pin
Number of channels
1
Type
RS-232, full duplex
Transmission rate [bit/s]
300 to 115K2 bit/s
Parity
None, odd, even
Data bits
7/8
Input signal level
Logic 1 (high) -3 to -15 V
Logic 0 (low) +3 to +15 V
Output signal level
Logic 1 (high) -5 to -15 V
Logic 0 (low) +5 to +15 V
Number of signals
Input: RxD, CTS
Output: TxD, RTS
Potential isolation
None
Cable type
Twisted pair, screened (e.g., LifYCY 3×2×0.20), 1 pair of wi-
res for ground (GND)
Process Controller Maxxis 5 PR 5900
4 Device installation
EN-48
Minebea Intec