2.4.3
Supply voltage connection
2.4.3.1
Version 230 V AC
Safe interruption of both supply voltage conductors must be
provided for, either by disconnecting the power connector or
using a separate switch.
The device is protected via two fuses (see Chapter
) on
the back of the device (primary side).
The device is equipped with a wide range power supply and
covers AC systems with a frequency of 50/60 Hz and a volta-
ge range of U
AC
= 100 to 240 V ±10% automatically (without
manual selection).
The power supply is protected against short circuits and
overloads, and disconnects automatically in case of a fault.
If the electrical protection has triggered:
-
Disconnect the device from all power sources and wait at
least 1 minute.
-
Determine and eliminate the cause of the error.
-
Reconnect the device to the supply voltage.
2.4.3.2
Version 24 V DC
This version is designed for U
DC
= 24 V.
The supply is established via two screw terminals (+/-).
The device is protected against incorrect polarity.
The device is protected in the + line via a fuse (see Chap-
ter
) on the back of the device (primary side).
2.4.4
Protective ground connection
2.4.4.1
Version 230 V AC
The device must be connected to a protective ground via a protective grounding
conductor (PE) in the network plug.
The power cable contains a protective grounding conductor which must not be
interrupted inside or outside the device.
The protective grounding conductor is connected to the back of the housing inside the
device.
Process Controller Maxxis 5 PR 5900