D-2
communication problem. Restart the pulse oximeter.
SpO2 Overrange
L
Measured SpO2 value is beyond the specified
measurement range. Check the animal’s condition.
PR Overrange
L
Measured PR value is beyond the specified measurement
range. Check the animal’s condition.
Sensor Off
L*
No Sensor
L
The SpO2 sensor detached the animal or the pulse
oximeter, or there was a fault with the SpO2 sensor, or an
unspecified SpO2 sensor was used. Check that the sensor
application site and the sensor type are correct, and make
sure that the sensor is undamaged. Reconnect the sensor if
the sensor is disconnected or use a new sensor if the
sensor is damaged.
SpO2 Low Perf
L
The signal detected is weak or the signal quality is poor.
Check the animal’s condition. Change the sensor
application site. If the problem persists, replace the sensor.
Voltage Too High
L
Voltage Too Low
L
The system power supply fails. Restart the pulse oximeter.
Battery Too Low
M
The battery power is low. Replace the batteries if alkaline
batteries are used or charge the battery if a lithium ion
battery is used.
Battery Error
L
A problem occurs when the lithium ion battery is being
charged. Check the battery for damage. If yes, replace the
battery.
Power Comm Err
M
Communication problem occurred to the power supply
part during the power-on self-test or operation. Restart the
pulse oximeter.
Clock Need Set
L
The real-time clock is reset. Set the clock.
Summary of Contents for PM-60Vet
Page 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 10: ...VIII FOR YOUR NOTES ...
Page 12: ...2 FOR YOUR NOTES ...
Page 30: ...2 8 FOR YOUR NOTES ...
Page 34: ...3 4 FOR YOUR NOTES ...
Page 56: ...6 6 FOR YOUR NOTES ...
Page 64: ...7 8 FOR YOUR NOTES ...
Page 78: ...10 2 FOR YOUR NOTES ...
Page 82: ...A 4 FOR YOUR NOTES ...
Page 96: ...P N 0852 20 77486 1 1 ...