5-6
WARNING
z
When the alarm sound is switched off, the pulse oximeter will give no
audible alarm tones even if a new alarm occurs. Therefore the user
should be very careful about whether to switch off the alarm sound or
not.
z
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjusting alarm volume to a low level may result in a hazard
to the patient. Always keep the patient under close surveillance.
5.4.3 Pausing the Alarm Tones
To pause the alarm tones, press the
key. In this case,
The audible alarm is paused, but the alarm lamp remains lit and the alarm
message remains displayed;
The remaining alarm pause time is displayed in the alarm status area;
The
symbol is displayed in the alarm status area.
Audible alarm starts again automatically after the alarm pause period expires. You can
also press the
key to restart the audible alarm. The audible alarm is reactivated
automatically when:
The pulse oximeter is shut down and restarted.
The pulse oximeter is switched to a new operating mode
The pulse oximeter enters or exits the standby mode
The pulse oximeter enters or exits the demo mode.
Alarm pause time can be adjusted and the default is 2 minute. To change the alarm
pause time,
Summary of Contents for PM-60Vet
Page 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 10: ...VIII FOR YOUR NOTES ...
Page 12: ...2 FOR YOUR NOTES ...
Page 30: ...2 8 FOR YOUR NOTES ...
Page 34: ...3 4 FOR YOUR NOTES ...
Page 56: ...6 6 FOR YOUR NOTES ...
Page 64: ...7 8 FOR YOUR NOTES ...
Page 78: ...10 2 FOR YOUR NOTES ...
Page 82: ...A 4 FOR YOUR NOTES ...
Page 96: ...P N 0852 20 77486 1 1 ...