6-5
To set the SpO
2
sensitivity,
1. Select
[
Menu
]
→
[
General Setup
].
2. Set
[
Sensitivity
] to [
High
], [
Med
] or [
Low
], whose averaging time is respectively
7 seconds, 9 seconds and 11 seconds.
6.5 Measurement Limitations
If you doubt the SpO
2
measurements, check the animal’s vital signs first. Then check the
pulse oximeter and SpO2 sensor. The following factors may influence the accuracy of
measurements:
Ambient light
Physical movement (animal or imposed motion)
Diagnostic testing
Low perfusion
Electromagnetic interference, such as MRI environment
Electrosurgical units
Dysfunctional haemoglobin, such as carboxyhemoglobin (COHb)and
methemoglobin (MetHb)
Presence of certain dyes, such as methylene and indigo carmine
Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2
Drop of arterial blood flow to unmeaurable level due to shock, anemia, low
temperature or vasoconstrictor.