5-7
1. Select
[
Menu
]
→
[
System
]
→
[
Maintenance >>
]
→
enter the required password.
2. Select
[
Alarm >>
] and then set [
Audio Paused
] to an appropriate time.
5.5 Setting the Alarm Level
The levels of all technical alarms except [
Sensor Off
] are predefined before the pulse
oximeter leaves the factory. To set the alarm level of [
Sensor Off
],
1. Select
[
Menu
]
→
[
Alarm Setup
];
2. Select
[
Sensor Off
] and choose a desired alarm level.
If you enter the demo mode after the [
Sensor Off
] alarm occurs, the alarm will be
presented in the form of prompt message when you exit the demo mode.
In the continuous monitoring mode, the alarm level of [
Sensor Off
] can be set to [
High
],
[
Med
] or [
Low
]. If theSpO
2
sensor falls off before a valid SpO
2
signal is obtained, the
pulse oximeter will present [
Sensor Off
] in the form of prompt message.
In the spot-check mode, the alarm level of [
Sensor Off
] can be set to [
High
], [
Med
],
[
Low
] or [
Off
].
If the [
Sensor Off
] alarm is not switched off, the pulse oximeter will present
the alarm in the form of prompt message in the case that the SpO
2
sensor falls
off before an effective SpO
2
signal is obtained.
If the pulse oximeter enters the standby mode after the [
Sensor Off
] alarm
occurs, it will present the alarm in the form of prompt message when the
standby mode is exited.
Summary of Contents for PM-60Vet
Page 1: ...PM 60Vet Pulse Oximeter Operator s Manual ...
Page 2: ......
Page 4: ...II FOR YOUR NOTES ...
Page 10: ...VIII FOR YOUR NOTES ...
Page 12: ...2 FOR YOUR NOTES ...
Page 30: ...2 8 FOR YOUR NOTES ...
Page 34: ...3 4 FOR YOUR NOTES ...
Page 56: ...6 6 FOR YOUR NOTES ...
Page 64: ...7 8 FOR YOUR NOTES ...
Page 78: ...10 2 FOR YOUR NOTES ...
Page 82: ...A 4 FOR YOUR NOTES ...
Page 96: ...P N 0852 20 77486 1 1 ...