7-4
Adjust [
Interval
]: The minimum interval is 2 seconds.
Select [
Start Time >>
] to set the [
Date
] and [
Time
] from which you
want to review.
Select [
Delete All
] to delete all the trend data under the current animal
ID.
Press the Right button to exit the trend window.
7.2.2 Spot-checking Mode
In the spot-checking mode, the pulse oximeter will stop measuring the animal if you
open the trend window. A animal cannot be admitted with the trend window opened.
To open the trend window, select [
Menu
]
→
[
Trend
].
In the trend window, SpO
2
and PR readings beyond the alarm limits are displayed in red.
If you have changed the system time before entering the trend window, the trend data
time before the system time is changed remains unchanged.
In the trend window, you can
Press the Up or Down button to page up or down, or press and hold the Up or
Down button for 1 second to speed paging up or down.