B-2
Table 2
Guidance and MINDRAY DS declaration
—
electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The
customer or the user of the equipment should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment
—
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
Transient/burst IEC
61000-4-4
Surge IEC
61000-4-5
Voltage dips,
Short interruptions
and voltage
variation on power
supply input lines
IEC 61000-4-11
Not applicable, the equipment is powered by internal battery.
Power Frequency
(50/60 HZ)
Magnetic Field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.