5-3
5.2 Messages Prompting
DPM1 can display various prompt messages. In Table 5-1, prompt messages as well as their
causes and solutions are listed.
Table 5-1 Table of Indications
Message
Cause
Solution
“Low Battery”
Batteries energy lower than 4.0
Voltage.
Replace batteries in time
“Memory Full”
The internal memory is almost full.
Stored data is to be covered. Export
data in time.
Blinking
“Memory Full”
Memory is full.
Stored data is being covered.
Export data in time.
“ID Full”
ID number Stored is greater than
95.
ID data is to be covered. Export
data in time.
Blinking “ID Full”
ID data is being covered.
Export data in time.
“Standby”
The device is in standby mode.
None.
“Communication”
The device is in communication
mode.
None.
DELETE ALL?
The Delete ID button is pressed.
Refer to 5.1.4 Data Deletion
ALL DELETED
The Delete ID button is pressed
again after “DELETE ALL?”
appears.
None.
DPM1 can also display technical error messages. In Table 5-2, error messages as well as
their causes and solutions are listed.
If the LCD screen can’t display anything, it may be damaged or error occurs during system
self-test. Please shut down the device (if can’t, remove the batteries) and contact our
Customer Service.
Table 5-2 Error indications
Error Message
Cause
Solution
Initiate Error
Failed self-test
Shut down the device (if can’t,
remove the batteries) and contact us
for service.
Please Release the
Button
Button error
Check for jammed button. If
problem remains, contact us for
service.
Pulse Not Found
Searching…
Pulse not found
Check the patient and alert the
doctor.