A-2
The DPM1 Pulse Oximeter with neonatal SpO2 sensors was also validated on adult subjects.
2.2 PR
Range:
25 to 254bpm
Resolution:
1bpm
Accuracy:
±
3bpm(For 520N)
±
2bpm(Others)
3. Display and Control
3.1 Display
Display Type:
Matrix LCD
Display Area:
Not less than 42mm
×
35mm.
Back Light:
Blue
Display Information:
SpO
2
, PR, Pulse strength, ID number, Memory Full, ID Full, Low
battery, Standby, Communication, Technical error
3.2 Buttons
Power Button:
Switches on/off the oximeter. The power-on is not delayed and the
power-off has a 2 second delay.
Backlight button:
Switches on/off the backlight
ID Confirm button:
Confirms whether use the previous ID for the new measurement.
Backlight ID
Confirm button
Deletes data
3.3 Power Saving Features
Refer to 5.3 Power Management
3.4 Connector
One dual-purpose socket for connecting SpO2 sensor and communication cable.
4. Electrical specifications
Working Voltage:
4.0 to 6.4 VDC
Power Supply:
Batteries
Battery Specifications:
Four Common 1.5V AA alkaline or rechargeable batteries
Shutdown Leakage Current:
< 200uA
Battery Run Time:
15-hour continuous operation with alkaline batteries
Power Consumption
720mW
5. Printing
Printer
The PC’s printer