3-5
Completed!
” is displayed.
6. Press
. The current screen exits.
7. After the accuracy calibration is completed, fill the 10ml, 20ml and 50ml
syringes with clean water respectively, and then set them according to the flow
rates and preset volumes in the table below. Take a clean empty container to
receive water, press
to start infusion. When the infusion volume reaches
the preset volume, the equipment triggers the [Complete] alarm, and the red
LED flashes. Then press
to stop the equipment. Put the water received in
the container onto an electronic balance with an accuracy of 0.1 g. The weight
of the received water should be within the standard range of the table below.
No.
Syringe
Specification
Flow Rate
Preset
Volume
Receiving
Standard
1
10ml syringe
100ml/h
5ml
4.90-5.10 g
2
20ml syringe
100ml/h
5ml
4.90-5.10 g
3
50ml syringe
500ml/h
25ml
24.50-25.50 g
Follow this procedure to perform an accuracy batch calibration:
1. Take a sample equipment that has been calibrated for accuracy, select [Main
Menu] → [Maintenance] → input the user maintenance password → [System
Calibration]→ [View Calibration Data], and obtain the accuracy calibration
parameters (Outer diameter, full scale and emptying) corresponding to brands
and specifications.
2. Take a sample equipment that has been calibrated for accuracy, and select
[Main Menu] → [Maintenance] → input the user maintenance password →
[System Calibration]→ [Input Calibration Data].
Align the front end of the sealing plug of
the push rod with the full scale line
Summary of Contents for BeneFusion SP3
Page 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Page 2: ......
Page 12: ...1 6 FOR YOUR NOTES ...
Page 24: ...2 12 FOR YOUR NOTES ...
Page 62: ...5 16 FOR YOUR NOTES ...
Page 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Page 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Page 78: ...7 8 FOR YOUR NOTES ...
Page 79: ......
Page 80: ...P N 046 013115 00 3 0 ...