5-13
5.2.6 Disassembling Rear Cover Assembly (Note: You should
wear an anti-static wristband throughout)
1. As shown in the figure: Remove the two M5x12 Phillips pan head screws that locks
the fastening clip assembly with a Phillips screwdriver.
2. As shown in the figure: Remove the fastening clip assembly.
3. As shown in the figure: Use the Phillips screwdriver to remove the two ST3x6 Phillips
pan head flat-end self-tapping screws that lock the power module.