4-4
replace the
control board.
SE121
/
AD Reference Voltage
Failure
The control board
is faulty.
Try to restart the
equipment. If it is
not recovered,
replace the
control board.
/
SA122
Charging Exception
1.Poor battery
connection.
2.The motor is
faulty.
3.The control
board is faulty.
1.Reconnect the
battery.
2.Replace the
battery with a
new one.
3.Replace the
control board.
SE201
/
Main and Auxiliary
Communication
Failure
1.The keypad
board is faulty.
2.The control
board is faulty.
1.Replace the
keypad board
2.Replace the
control board.
SE231
/
EEPROM Failure of
Control Board
The keypad
board is faulty.
Replace the
keypad board.
SE232
/
FLASH1 Failure of
Control Board
The keypad
board is faulty.
Replace the
keypad board.
SE234
/
Monitoring
CPU-EEPROM Read
and Write Errors
The control board
is faulty.
Replace the
control board.
SE251
/
Main control CPU
Self-Test Failure
The control board
is faulty.
Replace the
keypad board
SE252
/
Monitoring CPU
Self-Test Failure
The control board
is faulty.
Replace the
control board.
SE253
/
Main MCU Reset
Failure
The watchdog is
reset, there may
be software
defects.
Feedback to
R&D for
analysis.
SE254
/
Sub MCU Reset
Failure
The watchdog is
reset, there may
be software
defects.
Feedback to
R&D for
analysis.
SE257
/
Resource Failure
Software version
or resource file is
burnt incorrectly.
Re-burn the
software version
or resource file.
SE258
/
Brand Library Failure
Software version
or brand library is
burnt incorrectly.
Re-burn the
software version
or brand library.
/
SA260
Brand Library
Software version
Re-burn the
Summary of Contents for BeneFusion SP3
Page 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Page 2: ......
Page 12: ...1 6 FOR YOUR NOTES ...
Page 24: ...2 12 FOR YOUR NOTES ...
Page 62: ...5 16 FOR YOUR NOTES ...
Page 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Page 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Page 78: ...7 8 FOR YOUR NOTES ...
Page 79: ......
Page 80: ...P N 046 013115 00 3 0 ...