Ultrasonic milk analyser
Operation manual
11.04.19
47/110
Net number
0
- OK +
By using the button “up”▲ the operator has the possibility to increase the
number, showing the channel’s number, and by button “down”▼, to decrease
it. Pressing the button
Enter
saves the chosen channel and exits the function.
When connected in the production network each device has to have a unique
number.
7.2.3.2.Recalibrate.
Serves for changing definite calibration. Methods are described in point 6.4.
7.2.3.3. Save/Rest Cal.
Through this menu you may save the new calibration in the device or to
restore the old one (factory) calibration. This is necessary in case that you’ve
calibrated the device for cow milk, but after that the device is not measuring
correctly and you decide to restore the factory calibration settings. Position
the cursor across “
Restore calibration
” and press “
Enter
”
Possibilities:
Save calibration
– saves the chosen calibration in an internal buffer.
Restore calibration
– restores the chosen calibration from the internal
buffer.
The procedure
Save/Restore
is done for each calibration separately.
Current calibration content is not changed, the analyzer continues using it,
but there is a reserve copy in an internal buffer.
Current working
calibration (1,2 or 3)
Save
Internal storing
buffer
Current working
calibration (1,2 or 3)
Restore
Internal storing
buffer
Summary of Contents for LACTOSCAN SA
Page 12: ...Ultrasonic milk analyser Operation manual 11 04 19 12 110 Fig 3 Principle working scheme...
Page 13: ...Ultrasonic milk analyser Operation manual 11 04 19 13 110 Fig 4 Connecting peripheral devices...
Page 33: ...Ultrasonic milk analyser Operation manual 11 04 19 33 110 Fig 12 Peristaltic pump SL 16...
Page 93: ...Ultrasonic milk analyser Operation manual 11 04 19 93 110 Fig 14 Placing the probe...