39/110
your requirements you may make the calibration using a personal computer
type IBM PC and the company's calibration program or autonomous - by
recalibration.
When making corrections or calibrations be 100% sure in the accuracy of the
reference methods result.
6.4. Recalibrating the milk analyser
6.4.1. Running the analyser in mode Recalibrate
6.4.1.1.
Press the button
Enter
and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:
Release button
to start setup
After releasing the button on the display is shown:
Setup menu
Followed by the possible to be entered by the operator menus:
Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Accessories
Exit
6.4.1.3.
In the newly opened menu
,
position on
Recalibrate
and press the
button
Enter
.
Summary of Contents for LACTOSCAN SA