Ultrasonic milk analyser
Operation manual
11.04.19
44/110
2. Make sure it shows the same serial number as this already calibrated.
For checking, use the third sample with medium FAT content.
3. Measure the milk 5 times in the mode you’ve calibrated it.
In case that the device is not connected towards printer write down the
results.
4. Ignore the first two results.
The rest three could not differ more than 0,05% FAT, 0,07% SNF, 0,7%
Density one from another