Ultrasonic milk analyser
Operation manual
11.04.19
43/110
After 5th measurement completion automatically appears the menu, which
reminds us to place the
Low fat
milk sample:
Cal: Cow
Put Sampl: Low
5 times
Stir 2-3 times the milk sample before each measurement by pouring it from
one vessel to another. The needed quantity is poured in the sample-holder
and it is put in the recess of the analyser. Start the measurement by pressing
the button
Enter
. The sample is sucked. Appears the following menu:
Make 5 times measurement of the low FAT sample.
After 5th measurement completion automatically appears the menu:
Cal: Cow
Put sample: Water
5 times
Which reminds for 5-times water measurement.
After the 5
th
measurement appears the menu:
Recalibrated
Power Off-On
This means that the calibration was completed successfully and the analyzer
is recalibrated for cow milk, marked as “Cal: Cow”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.
Next time when the analyser is switched on, it will be ready for work with
those milk types it was just calibrated with.
If calibration with another milk type is needed, do not forget to change the
calibration number for the new type of milk.
Calibration for Sheep milk will be saved as second calibration, UHT – as
third. This sequence may not be followed and calibrations can be saved in
whichever order is needed. Calibration can be done with different liquid dairy
products using 2 representative samples.
Checking the calibration
1. Switch on the calibrated device.
Summary of Contents for LACTOSCAN SA
Page 12: ...Ultrasonic milk analyser Operation manual 11 04 19 12 110 Fig 3 Principle working scheme...
Page 13: ...Ultrasonic milk analyser Operation manual 11 04 19 13 110 Fig 4 Connecting peripheral devices...
Page 33: ...Ultrasonic milk analyser Operation manual 11 04 19 33 110 Fig 12 Peristaltic pump SL 16...
Page 93: ...Ultrasonic milk analyser Operation manual 11 04 19 93 110 Fig 14 Placing the probe...