Immerse the electrode far enough into the solution to insure the reference
junction is below the surface.
Allow adequate time for the electrode to stabilize in standards and samples
before taking a reading.
Clean the electrode periodically. Allow more time for aged electrodes.
Do not use the pH electrode in solutions of fluoride ion at low pH. This will
etch the glass membrane.
Sulphide vapors can permeate the electrode wick and contaminate the
reference element. Minimize contact in such environments and change the
reference electrolyte frequently.
6.Preparation for pH measurement
When the analyzer is with pH measuring option, it is received from the
customer with pH probe packed separately and there’s a stopper on its place.
If you need to measure pH follow the procedure below:
1. Loose the nut anti-clockwise.
2. pull up the stopper