Ultrasonic milk analyser
Operation manual
11.04.19
41/110
With the button ▲(
Next
), choose
SNF
and in the same as above
described procedure, enter the value for
SNF
. After it is finished, press
“
Enter
” (OK) and you are going back to the previous menu. With button
▲ (
Next)
, choose
DEN (density
) and enter the value for density
; the rest
of the parameters are entered in the same manner – LAC (lactose), SOL
(salts), PRO (protein)
Cal1 Cow
High
PRO=f.ff
Edit OK Next
If you miss to enter some of the parameters of milk, the following warning
message will appear:
You Must Enter
Values > 00.00
Try Again
Then you must press the button
Enter
(ОК)
and enter the missed parameters.
After all the parameters are entered, press
Enter
(ОК).
You must enter values for all the measured milk parameters!!!
The screen for entering the results, received with the corresponding
reference methods (See
Appendix Methods)
for the
low-fat
milk is displayed:
Cal1 Cow
Low
FAT=f.ff
Edit OK Next
In the same way the values of milk with low fat sample are entered.
You must enter values for all the measured milk parameters!!!