Appendix B
DVVA II 4000
B-2
400-42803-01 (Rev B) - July 2017
Blank Correction
A blank analysis is performed to correct for the frictional force between the piston and the sample
chamber, and for the deflection of the linear transducer.
where
F = applied force
F
S
= force on sample
F
f
= frictional force
h
b
= bed height
h
= measured height
h
d
= deflection
Void Volume
where
V
A
= apparent volume
F
s
= force of sample
h
b
= bed height
m
= mass
S
= skeletal density
d
= diameter of sample chamber
F
S
F
F
f
h
b
–
=
h
b
h
h
d
F
–
=
VV F
S
V
A
F
S
V
T
–
100
=
V
A
F
S
h
b
A
=
A
d
2
4
---------
=
V
T
m
S
-----
=