Appendix A
DVVA II 4000
A-8
400-42803-01 (Rev B) - July 2017
2486- Cannot construct (name) report type. Program will terminate.
2487-
Cannot start report generator. Error code (number). Program will terminate.
2488-
File (file name) cannot be opened for editing. It is already in use.
2489-
File (file name) cannot be opened for writing. It is already in use.
2490-
No ‘.INI’ file present. Application will terminate.
Cause A:
You may not have full rights to the applications’s folders and files.
Cause A:
You may not have full rights to the applications’s folders and files.
Action B:
An internal processing and/or hardware error has occurred.
Action A:
Contact your service representative if you continue to receive this
error.
Cause:
The file you specified is being used in another edit operation.
Action:
Check the Windows list to locate the other edit session.
Cause:
The file you specified in a
Save As
operation is already open for edit.
Action:
Select a different file for the
Save As
operation.
Cause:
The ASCII (.INI) file containing initialization and system options
information cannot be found. The .INI file may have become cor-
rupted. The application cannot operate without this file.
Action:
Use the Set Up CD to uninstall the analysis program. When you unin-
stall the program, only application files are deleted; data files remain
intact.
After the uninstall operation is complete, reinstall the program.
Summary of Contents for DVVA II 4000
Page 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Page 4: ......
Page 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Page 16: ......
Page 32: ......
Page 116: ......
Page 136: ......
Page 160: ......
Page 180: ......
Page 212: ......
Page 214: ......
Page 240: ......
Page 246: ......
Page 248: ......