DVVA II 4000
Clean System
400-42803-01 (Rev B) - July 2017
6-9
Clean System
Cleans residual sample from the inner walls of the sample chamber; the Clean System dialog is
displayed.
1. Wipe off the upper piston tip.
2. Remove the sample tray and insert the waste tray.
3. Press the
Reset
button on the instrument’s front panel.
4. Click
Start
on the Clean System dialog.
5. When the cleaning process is complete, all components return to a ready state and the dialog
closes automatically.
6. Remove the waste try and empty its contents.
7. Wipe off the lower piston tip.
8. Insert the sample tray.
9. Press the
Reset
button on the instrument’s front panel.
Start
Starts the cleaning process. The piston travels down the
sample chamber, removing any residual sample as it
travels.
Cancel
Cancels the operation; all components return to a ready
state.