DVVA II 4000
Troubleshooting
400-42803-01 (Rev B) - July 2017
9-9
2. If the safety shield is not already raised, raise it to its upper position.
3. Retrieve the pellet removal tool (shown below) that was included in the instrument
accessories kit.
4. Carefully place the tool into the compression chamber and while applying pressure, rotate
the tool.