ASAP 2460 Operator’s Manual
Appendix B
246-42800-01 - Dec 2013
B-43
6522-
Time limit exceeded while evacuating manifold.
Cause A:
Maximum manifold evacuation time was exceeded before the vacuum
set point was achieved. Vacuum pump may be turned off.
Action A:
Turn on vacuum pump switch. Then restart the analysis.
Cause B:
The vacuum pump oil level is low.
Action B:
Check the vacuum pump oil level and add more oil if necessary. Then
restart the analysis.
Cause C:
The manifold is contaminated or leaking.
Action C:
Correct the problem. Refer to Appendix D,
6523-
Analysis suspended: Time limit exceeded while evacuating sample
(unrestricted).
Cause:
The maximum time for evacuating the sample through the unrestricted
valve was exceeded.
Action:
Check the sample tube and the sample tube fitting; ensure that the tube
is securely attached to the port. Verify that the sample is properly
degassed; then restart the analysis.
Action A:
Set the analysis gas regulator to 15 psig (1.0 bar). Then resume the
analysis.
Cause B:
The analysis gas bottle is empty.
Action B:
Connect a new analysis gas bottle. Then resume the analysis.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...