ASAP 2460 Operator’s Manual
Performing an Analysis
246-42800-01 - Aug 2013
2-39
Performing an Analysis
Analyses may be performed in one of these modes:
• Standard
• Krypton
• High Throughput
• Micropore
Ensure that the analysis gas and the Psat gas (if used) specified in the sample file match the system
configuration gases; if they do not, correct the sample information file or correct the unit configuration
gas. You can check the system configuration gas by selecting
Unit > Unit Configuration
(refer to
Standard Analysis
Use the following guidelines for performing standard analyses:
• Standard analyses cannot be performed if a high throughput or krypton analysis is in progress.
• One analysis can be started at a time.
• A sample can be added to any idle port and an analysis started without disturbing the analyses being
performed on other ports.
• Samples can be removed from any of the six ports without disturbing the analyses being performed
on other ports.
• The sample dosing method should be
Normal
.
• All analyses must use the same gas.
• If saturation pressure is being measured, all analyses must use the same Psat gas.
A manual evacuation can be performed on any idle port while analyses are in progress.
Before beginning an analysis, make sure the tank pressure for the gas regulator is at
least 200 psig. Pressures less than 200 psig may cause the sample to be
inadequately saturated, resulting in inaccurate data or termination of analysis.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...