Appendix B
ASAP 2460 Operator’s Manual
B-22
246-42800-01 - Dec 2013
4031-
Not enough points with a relative pressure in the range (n,n)
Cause:
Fewer than two data points selected for the Dubinin report falls within
the selected relative pressure range.
Action:
Edit the calculation assignments or the fitted relative pressure range.
4033-
Not enough points to generate Dubinin Tabular Report.
Cause:
There are fewer than two valid data points available for the Dubinin
tabular reports.
Action:
Examine the calculation assignments in the Collected Data dialog of
the sample file. You must have at least two micropore pressures
selected for inclusion in the Dubinin report.
4034-
Fewer than 2 points available for Dubinin calculations.
Cause:
There are fewer than two valid data points available for Dubinin
reports in one of the sample files selected for overlaying.
Action:
Examine the calculation assignments in the Collected Data dialog of
the sample file. You must have at least two micropore pressures
selected for inclusion in the Dubinin report.
4035-
Cannot calculate optimized Astakhov exponent: Not enough points with a rel-
ative pressure in the range [(pressure), (pressure)].
Cause:
There are fewer than two valid data points in the relative pressure
range specified. Astakhov reports will not be produced.
Action:
Examine the calculation assignments in the Collected Data dialog for
the sample file. You must have at least two pressures in the given range
selected for inclusion in the Dubinin report.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...