Analysis Conditions Files
ASAP 2460 Operator’s Manual
3-22
246-42800-01 - Aug 2013
Dose Increments
option (continued)
•
Low Pressure incremental dose mode
- Choose this option when you
are performing an analysis of microporous materials. At low pressures
on Type 1 isotherms, the pressure points are very closely spaced, making
a useful pressure table difficult to define. Choosing this mode enables
you to measure equilibrium points at approximately equal intervals on
the quantity adsorbed axis. Each dose is fully equilibrated and recorded
as a data point.
In this mode, the sample is successively dosed with a specified amount
of gas until the first pressure point is reached. The first point is the
threshold value, triggering the transition from Incremental Dose Mode
to Pressure Table Mode. When the first pressure table value is reached,
Incremental Dose Mode is disabled, and points are recorded in
accordance with the specified pressure table. Because the data points
recorded during Incremental Dose Mode may define most of the
analysis, one point on the pressure table can be sufficient and serve as
the end point for the analysis.
This field is disabled if
First Pressure Fixed Dose Mode
is selected.
•
Dose amount
- Enter the amount of gas to be added to the sample for
each data point until the first point on the pressure table is reached.
Enabled when
Low pressure incremental dose mode
is selected.
•
Equilibration delay
- Enabled when
Low pressure incremental dose
mode
is selected.
Minimum:
Prevents premature equilibration caused by reduced
percentage sensitivity to pressure changes at the lowest pressures
.
Maximum:
Prevents the effects of long term temperature or pressure
drift, which may cause the instrument to wait an excessive length of time
for equilibration.
Field or Button
Description (continued)
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...