8 Error Messages
4412
Error calibrating the servo.
Cause:
Calibration results are out of range.
Action:
Follow standard calibration procedures and try again. If the problem persists, contact
your Micromeritics service representative.
4413
Error overheating on port [0]. Current = [TEMPC0] [TEMPC-U], Target
= [TEMPC0] [TEMPC-U], Limit = [TEMPC0] [TEMPC-U].
Cause:
The temperature of the indicated mantle exceeded the maximum allowed value.
Action:
Ensure the power and thermocouple connectors for the mantle are properly installed. If
they are, the mantle may be defective; contact your Micromeritics service rep-
resentative.
4414
Error thermocouple unplugged on port [0]. Target = [TEMPC0]
[TEMPC-U]."
Cause:
The thermocouple is unplugged or has malfunctioned.
Action:
Ensure the thermocouple is plugged in. If the problem persists, contact your Micro-
meritics service representative.
4420
Communications error.
Cause:
The application failed to connect to the Smart VacPrep.
Action:
Ensure the unit is powered and properly connected to the network specified during install-
ation. Check the power cable, power switch, and Ethernet cable, then reconnect to the
Smart VacPrep either through the Smart VacPrep menu for this instrument or by restart-
ing the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4421
Smart VacPrep S/N %s is busy and could not be removed.
Cause:
The Smart VacPrep cannot be removed because it is currently performing an operation.
Action:
Wait until the Smart VacPrep completes the current operation and try again.
8 - 2
Smart VacPrep Operator Manual
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...