1 About the Smart VacPrep
l
QuickStart
menu option and Smart VacPrep panel buttons. Accessed from the
Smart VacPrep >
Unit [n] > QuickStart Degas Conditions
menu selection. Provides the option to enter commonly used
degas conditions for each Smart VacPrep port without the need for a sample information file.
This option provides three ways to load information into
QuickStart Degas Conditions
:
1. Load degas conditions from an existing degas conditions file
2. Copy conditions from another Smart VacPrep unit and port
3. Enter degas conditions manually
When using the
QuickStart
menu option, the degas process is started using the
Select
button and
Start
buttons for each port on the Smart VacPrep unit.
Using the
Start Degas
option does not change the degas conditions set up in
QuickStart
.
QuickStart
degas conditions are remembered for the next push button operation.
Sample tubes are attached to the sample ports. Heating mantles are then installed and secured to the
sample tube with a metal clip. If using straight wall sample tubes, heating mantles with elastic cords
may be purchased separately.The mantle is plugged into the power and thermocouple connectors on the
front panel. The degassing process can then be started either by pressing the
Select
and
Start
buttons on
the Smart VacPrep unit or by going to
Smart VacPrep > Unit [n] > Degas > Start Degas
.
A Check Seal or TransSeal can be used to transfer air-sensitive samples from the Smart VacPrep to the
analyzer's port without atmospheric contamination. To order these parts, reference the Parts and
Accessories section for the analyzer. Instructions for using the Check Seal or TranSeal are included
with the ordered part.
Component
Description
Cancel button
Stops the degas process on all ports where the
Select
indicator is lit.
Ethernet port
Located on the left side panel. Use to connect an Ethernet switch.
Power Switch (not shown)
Located on the back of the unit, turns the unit on or off.
Select buttons
Used to select the ports to be degassed.
Start button
Starts the degas process on all ports where the
Select
indicator is lit.
Status Indicator
Three lights indicate the current status of the station. The lights flash
when the valves are open.
Green
. Indicates the station is idle and not hot. The setpoint temperature
is below 30 ºC and the actual temperature is below 50 ºC.
Red
. Indicates when the setpoint temperature is greater than 30 ºC and
heat is being applied to the station.
Yellow
. Indicates the station is in use, but not currently being heated. The
heating mantle may still be above 50 ºC.
Smart VacPrep Components Table
1 - 2
Smart VacPrep Operator Manual
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...