Preventive Maintenance
ASAP 2460 Operator’s Manual
7-18
246-42800-01 - Aug 2013
8.
Close valves 8 and 18, and record the pressure as the
Initial Reading
for Inlet Ports
in the table
below.
9.
Wait 3 minutes, then record the pressure as the
3 Minute Reading
in the table above.
10. Subtract the first reading from the second reading and record in the
Difference
column in the
table above.
11. If the value in the
Difference
column is at or below the value in the
Limits
column, enter Yes in
the
OK?
column. If the Difference value is not below the Limits value, a gas inlet valve, inlet
plug, or gas line is leaking from atmosphere.
12. Close all gas inlet manifold valves (21 through 26).
13. Record the pressure as the
Initial Reading
, then begin timing as soon as you complete the next
step.
14. Gas inlet valves 21 through 26 must remain closed during this procedure. Pressurize the inlet to
valve 21 by opening the supply valve or removing the port plug. This allows gas or air to
pressurize the inlet valve above the seat.
15. After 3 minutes, record the pressure as the
3 Minute Reading
. Subtract the first reading from the
second and record in the
Difference
column in the table above.
16. Repeat steps 13 through 15 for the inlet valves 22 through 26.
Test
Initial Reading
3 Min. Reading
Difference
Limits
OK?
Inlet Ports
30 microns
(0.3 mmHg)
Valve 21
2.1 microns
(0.0021 mmHg)
Valve 22
2.1 microns
(0.0021 mmHg))
Valve 23
2.1 microns
(0.0021 mmHg)
Valve 24
2.1 microns
(0.0021 mmHg)
Valve 25
2.1 microns
(0.0021 mmHg)
Valve 26
2.1 microns
(0.0021 mmHg)
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...