Cleaning the cuff
You can machine wash the
cuff cover
at 30°C (do not iron!).
WARNING:
Under no circumstances, however, may you
wash the inner bladder! Always remove the sensitive
bladder from the sleeve before washing and replace it care-
fully again afterwards.
Accuracy test
We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
Disposal
13. Guarantee
This instrument is covered by a
3 year guarantee
from the date
of purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) confirming date
of purchase or the till receipt.
Please contact Microlife-Service (see foreword).
Batteries and electronic instruments must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
x
The guarantee covers both instrument and cuff. Batteries are
not included.
x
Opening or altering the instrument invalidates the guarantee.
x
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
Accuracy test
We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
contact aponorm
®
by microlife-Service to arrange the test (see
foreword).
5
Please contact aponorm
®
by microlife-Service (see foreword).
15.
14. Technical Specifications
Technical alterations reserved!
Operating temperature:
10 - 40 °C / 50 - 104 °F
Storage temperature:
-20 - +50 °C / -4 - +122 °F
15 - 90 % relative maximum humidity
Weight:
735 g (including batteries)
Dimensions:
160 x 140 x 98 mm
Measuring procedure:
oscillometric, corresponding to Korot-
koff method: Phase I systolic, Phase V
diastolic
Measurement range:
30 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
x
4 x 1.5 V Batteries; size AA
x
Mains adapter DC 6V, 600 mA
(optional)
Reference to
standards:
EU Directives 93/42/EEC
NIBP requirements: EN 1060-1 /-3 /-4,
ANSI / AAMI SP10
16.