[Measuring program
name]: The object can't be
fitted (Optimized fitting
failed)
The target object could not be
calculated in the measuring
program specified, as the opti-
mized fit has failed.
Check the threshold for the opti-
mized fit in the measuring program
specified or deactivate optimized fit.
[Measuring program
name]: The object can't be
fitted (max iteration
achieved)
The target object could not be
calculated in the measuring
program specified, as the maxi-
mum number of iterations has
been reached.
Check the parameters of the meas-
uring program specified.
[Measuring program
name]: The object can't be
fitted
The target object could not be
calculated in the measuring
program specified.
Check the parameters of the meas-
uring program specified.
[Measuring program
name]: Input object isn‘t
supported
The specified measuring pro-
gram does not support the type
of input object selected.
Check the configuration of the
measuring program input data (see
Chapter
[Measuring program
name]: Input objects aren't
supported in this combina-
tion
The specified measuring pro-
gram does not support the
combination of input objects se-
lected.
Check the configuration of the
measuring program input data (see
Chapter
[Measuring program
name]: Reference plane
doesn't intersect given ROI
The specified measuring pro-
gram could not determine the
measured values correctly, as
the reference plane does not in-
tersect the ROI specified in the
measuring program.
in the measuring program specified
or check the parameters of the
measuring program generating the
reference plane (see Chap.
[Measuring program
name]: Structured point
cloud is required
Internal error.
Contact Micro-Epsilon.
[Measuring program
name]: Mesh information is
required
Internal error.
Contact Micro-Epsilon.
[Measuring program
name]: Structured point
cloud or mesh information
are required
Internal error.
Contact Micro-Epsilon.
[Measuring program
name]: Invalid port index
Internal error.
Contact Micro-Epsilon.
Input/output: Error meas-
ured values output
An error occurred during output
of the measured values via
Ethernet.
Check the Ethernet connection and
Ethernet interface used for output-
ting the measured values. Check
the settings for the output of meas-
ured values (see Chap.
Input/output: Error in-
put/output interface
An error occurred during initiali-
zation of the Ethernet interface
for outputting the measured val-
ues.
Check the Ethernet interface used
for outputting the measured values.
Check the settings for the output of
measured values (see Chap.
Ready
No error. A point cloud was
loaded (offline mode).
-
The file [...] could not be
opened!
The specified name or path of
the point cloud is incorrect.
Select the correct directory contain-
ing your captured point cloud.
Summary of Contents for reflectCONTROL RCS
Page 1: ...3DInspect 1 1 Instruction Manual ...
Page 25: ...Working with 3DInspect Page 25 ...
Page 72: ...Description of the Data Preprocessing View Page 72 ...
Page 74: ...Description of the Data Preprocessing View Page 74 ...
Page 129: ...Appendix Page 129 ...
Page 133: ......