Page 21
Installation and Assembly
optoCONTROL CLS1000
Switching output Q1 properties
Sensor
pin
PC1000-x
cable color
Comments
V
+
max. 12 ... 30 VDC, I
max
= 100 mA
Signal level depends on supply voltage
Switching behavior programmable: NPN, PNP, PP
4
Black
Switching output Q1
3
Blue
Ground
4.5.3.4 Trigger Input CLS1000-AU-xx-T
Only applies for controller CLS1000-AU-xx-T with trigger function.
Trigger
1
5
4
Q
PP / PNP / NPN
0 V
10 V
2
GND
3
U Analog
V
+
Brown
White
Black
Gray
Blue
Fig. 13 Circuit for CLS1000-AU-xx-T
Trigger input properties
Sensor
pin
PC1000-x
cable color
Comments
3
Blue
Ground
max. 30 VDC; input current
I
max
= < 1 mA
5
Gray
Trigger (IN)
When wiring the trigger input with an NPN output, an external resistance
≥ 4.7 kΩ ≤ 10 kΩ is necessary.