Installation and Assembly
optoCONTROL CLS1000
4.
Installation and Assembly
4.1
Installation and Mounting
i
Ensure careful handling during installation and operation.
During installation of the optoCONTROL CLS1000, ensure that the sensors can be freely moved and
are not exposed to any pronounced curvatures or sharp corners. Damage to or destruction and/or
failure of the sensor and/or influence on the test result.
The smallest bending radius of the sensor (optical fiber) is 18 mm (3 x Ø cable diameter).
M18x1
M12x1
106 (4.17) 85 (3.25) 75 (2.95)
32 (1.26)
48.9 (1.93)
14.5 (0.57)
ø4.3
42.4 (1.67)
28.4 (1.12)
28 (1.10)
18.5 (0.73)
Fig. 4 Dimensional drawing of optoCONTROL CLS1000, dimensions in mm (inches)
Do not allow the end surfaces of the sensor to hit against edges or surfaces.
> Reduced signal quality or failure of the measuring device.
Mount the controller exclusively on a flat surface or a DIN rail using the existing through-holes on the housing
.
Any type of clamping is not permitted.
Connect the controller according to the pin assignment