3.12 High pressure pump
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
1
2
3
Figure 22
Mixing coil for low pressure gradients connection
1
Inlet valve holder
2
PEEK pressure screw, short 6.2744.070
3
Mixing coil for low pressure gradients
6.2758.020
3.12.3
Deaerating the high pressure pump
The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.
CAUTION
The high pressure pump must
not
be deaerated before the initial start-
up
Deaerate the high pressure pump as follows
Deaerating the high pressure pump
The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.