
3.2 Initial installation
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
2 Installations on the rear of the instrument
■
.
■
Place the detector in the instrument and connect it
■
Remove transport locking screws
(see "Removing transport lock-
.
■
Connect the leak sensor
(see "Connecting the leak sensor", page
.
■
Connect drainage tubings
(see "Installing drainage tubing", page
.
3 Installing the eluent path
■
Equip the eluent aspiration tubing and connect with the eluent
bottle
.
■
Instead of the column, connect the 6.2744.040 coupling with a
6.2744.010 pressure screw at the end of the pre-installed column
inlet capillary.
■
Use a 6.2744.014 pressure screw to connect the capillary of the
MSM-HC labeled with
Eluent
to the other end of the 6.2744.040
coupling.
■
Use a 6.2744.090 long pressure screw to connect the capillary of
the MSM-HC labeled with
Detector
to the input of the MCS
■
Connect the detector inlet capillary with a 6.2744.090 long pres-
sure screw to the output of the MCS
.
4 Installing the sample path
NOTE
The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.
■
Connect the 6.1803.040 sample aspirating capillary
ted at the sample input of the injection valve to the output of the
sample degasser with the aid of a 6.2744.090 long pressure
screw
■
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through.