3.19 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
60
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
3 Connecting the pressure side
NOTE
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A:
a 6.2744.180 pump tubing connection
with filter
or
■
Case B:
a 6.2744.160 pump tubing connection
without filter
.
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection
with
filter
must
be used.
Case A:
6.2744.180 pump tubing connection with filter:
1
2
3
Figure 33
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
Slide union nut
onto the pump tubing.
■
Select a suitable adapter
and slide it onto the pump tubing.
The type of adapter depends on the pump tubing
■
Place the tubing olive with filter holder
onto the pump tub-
ing.
■
Screw the union nut
.
or
Case B:
6.2744.160 pump tubing connection without filter: