3.9 Eluent
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■
Fasten the bottle attachment
The aspiration filter
must rest on the base of the eluent
bottle.
6 Mounting the adsorber tube
NOTE
In the case of alkaline eluents and eluents with lower buffer
capacity, the eluent bottle must be equipped with a CO
2
adsorber
.
■
First, place a piece of wadding
, then the CO
2
adsorber
in the large opening of the adsorber tube
with the plastic cover.
■
Fasten the adsorber tube
the bottle attachment
.
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 16
Eluent bottle – connected
1
Eluent aspiration tubing 6.1834.080
For aspirating the eluent. Pre-installed.
2
Adsorber tube 6.1609.000
3
Wadding
4
CO
2
adsorber
Adsorbs CO
2
from the air (e.g. Merck soda
lime with indicator, no. 6839.10).