
6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
110
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
Problem
Cause
Remedy
MCS – vacuum pump
defective.
Contact the Metrohm Service.
Background conduc-
tivity too high
MSM-HC – not connected.
MCS – not connected.
Connect the MCS.
Incorrect eluent.
(see Chapter 5.4.2.3, page 80)
MSM-HC – regeneration or
rinsing solution flow prob-
lems.
Check the flow of the regeneration and rinsing
solution
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than
expected
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
MSM-HC – No (or
insufficient) pump-
ing of regeneration
or rinsing solution
Leak in the system.
Check connections.
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure .
Peristaltic pump – filter
blocked .
Replace the filter .
MSM-HC – backpressure
too high.
.
Peristaltic pump – pump
tubing defective.
Replace pump tubing .
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column
.