3 Installation
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
57
Connecting the capillaries of the MSM-HC
1 Connecting the eluent inlet capillary
■
Fasten the end of the inlet capillary labeled with
Eluent
with a
6.2744.070 short PEEK pressure screw to the output of the col-
umn.
2 Connecting the eluent outlet capillary
■
Fasten the end of the outlet capillary labeled with
Detector
with a
long PEEK pressure screw (6.2744.090) to the input of the MCS.
OR
Connect the end of the outlet capillary labeled with
Detector
and
the detector inlet capillary with a 6.2744.040 coupling and two
6.2744.070 short pressure screws.
3 Connecting the rinsing solution inlet capillary
■
Fasten the end of the inlet capillary labeled with
H20
with a
6.2744.070 short PEEK pressure screw to the pump tubing con-
nection of the pump tubing which carries the rinsing solution.
4 Connecting the rinsing the solution outlet capillary
■
Guide the other end of the outlet capillary labeled with
Waste
into a sufficiently large waste container and fasten it there.
5 Connecting the regeneration solution inlet capillary
■
Fasten the end of the inlet capillary labeled with
H2S04
with a
6.2744.070 short PEEK pressure screw to the pump tubing con-
nection of the pump tubing which carries the regeneration solu-
tion.
6 Connecting the regeneration solution outlet capillary
■
Guide the other end of the outlet capillary labeled with
Waste
into a sufficiently large waste container and fasten it there.
The rinsing and regeneration solutions are pumped with a peristaltic pump