5.11 High Capacity Metrohm Suppressor Module (MSM-HC)
■■■■■■■■■■■■■■■■■■■■■■
100
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
5.11.3.3
Replacing parts of the MSM-HC
It may be necessary to replace parts of the MSM-HC in the following
cases:
■
Loss of suppression capacity which cannot be eliminated (reduced
phosphate sensitivity and/or significant rise in the baseline).
■
Blockage of the MSM-HC which cannot be eliminated (solutions can
no longer be pumped through the suppressor).
and the MSM-HC connecting piece
with its supply and discharge lines can be replaced.
Replacing parts of the MSM-HC
Replace parts of the MSM-HC as follows
1 Disconnecting the MSM-HC from the IC system
■
Switch off the instrument.
■
Disconnect the MSM-HC from the separation column, peristaltic
pump and detector.
2 Dismantling the MSM-HC
■
■
Pull the MSM-HC connecting piece
. The connecting
piece and the MSM-HC rotor A normally stick to one another – if
this is not the case: Take a sharp object, insert into the slot
in the MSM-HC housing, and pull out the MSM-HC rotor A
in this way.
■
Detach the MSM-HC connecting piece
3 Cleaning the new MSM-HC rotor A
■
Clean the sealing surface of the new MSM-rotor A
ethanol using a lint-free cloth.
4 Inserting the new MSM-HC rotor A
CAUTION
An incorrectly inserted MSM rotor A
can be
destroyed
during start-up.