6 Troubleshooting
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
109
Problem
Cause
Remedy
Eluent – Evaporation of the
organic solvent in the elu-
ent.
Check the eluent bottle attachment
.
Peak areas lower
than expected
Sample – Leak in the sam-
ple path.
Check the sample path.
Sample – Blockage in the
sample path.
Check the sample path.
Sample – Sample loop not
(completely) filled.
Prolong the sample transfer time.
Sample – Gas bubbles in
the sample.
Use the sample degasser
(if available).
MCS – not connected.
Connect the MCS.
Peristaltic pump –
insufficient or no
delivery rate
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure
Peristaltic pump – filter
blocked.
.
Peristaltic pump – pump
tubing defective.
.
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves
valve and outlet valve", page 86)
.
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
High pressure pump –
defective piston seals.
.
MCS –
CO
2
adsorption cartridge
exhausted.
Replace CO
2
adsorption cartridge
Pulsation damper not con-
nected.
Pulsation damper not con-
nected.
Connect pulsation damper
.