3 Installation
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
45
5
1
2
6
7
5
3
4
5
Figure 23
Deaerating the high pressure pump
1
Syringe 10 mL 6.2816.020
For aspirating the eluent.
2
Luer connector
On purging needle.
3
Purging needle 6.2816.040
4
Deaerating capillary
5
PEEK pressure screws, short
6.2744.070
6
Purge valve
7
Purge valve rotary knob
1 Connecting the purging needle
■
Push the end of the purging needle
over the end of the
deaerating capillary
on the purge valve.
2 Connecting the syringe
■
of the purging
.
3 Opening purge valve
■
½
rotation counterclock-
wise.
4 Setting the flow rate
■
Start MagIC Net (if not yet started).
■
Ensure that the eluent aspiration tubing is immersed sufficiently in
the eluent.
■
Let the high pressure pump run.