46
Figure 54: N-Cycle Burst Interface (Page 2/2)
Function Menu
Settings
Explanations
Trig Delay
Set the delay time before the burst starts.
Trig Out
Off
Disable trigger out.
On
Enable trigger out.
Page 2/2
Return to the previous page.
Table Menu Explanations of the N-Cycle Burst (Page 2/2)
Infinite
In infinite mode, the cycle number of the waveform is set as an infinite value. The generator outputs a continuous waveform after
receiving the trigger signal. Waveforms that support infinite mode include sine, square, ramp, pulse and arbitrary.
Press
[Burst] → [NCycle] → [Infinite]
, and set the trigger source to external or manual. The screen will display an infinite cycle
burst, as shown in Figure 55.
Figure 55: Infinite Burst Interface
Gated
In gated mode, the generator controls the waveform output according to the gate signal level. When the gated signal is true the
generator outputs a continuous waveform. When the gated signal is false the generator first completes the output of the current
period and then stops. Waveforms that support gated burst include sine, square, ramp, pulse, noise and arbitrary.
Press
[Burst] → [Gated]
, to enter the following interface.