43
Function Menu
Settings
Explanations
Sweep time
Set the time span of the sweep in which the frequency changes from the start frequency to stop
frequency.
Start Freq
Mid Freq
Set the start frequency of the sweep.
Set the center frequency of the sweep.
Stop Freq
Freq Span
Set the stop frequency of the sweep.
Set the frequency span of the sweep.
Source
Internal
Choose internal source as a trigger.
External
Choose external source as a trigger. Use the [Aux In/Out] connector at the rear panel.
Manual
Trigger a sweep manually.
Trig Out
Off
Disable trigger out.
On
Enable trigger out.
Page 1/2
Enter the next page.
Table Menu Explanations of Sweep (Page 1/2)
Figure 50: Setting Interface of Sweep (Page 2/2)
Function Menu
Settings
Explanations
Sweep time type
Linear
Set the sweep with linear profile.
Log
Set the sweep with logarithmic profile.
Direction
Up
Sweep upward.
Down
Sweep downward.
Page 2/2
Return to the previous page.
Table Menu Explanations of Sweep (Page 2/2)
Sweep Frequency
Use start freq and stop freq or center freq and freq span to set the range of the frequency sweep. Press the key again to switch
between the two sweep range modes.
Start Frequency and Stop Frequency
Start Frequency and Stop Frequency are the lower and upper limits of the frequency for sweep. Start Frequency ≤ Stop Frequency.
Choose
[Direction] → [Up]
, the generator will sweep from Start frequency to Stop frequency.
Choose
[Direction] → [Down]
, the generator will sweep from Stop frequency to Start frequency.
Center Frequency and Frequency Span
Center Frequency = (|Start Fre Stop Frequency|)/2Frequency
Span = Stop Frequency - Start Frequency
Sweep type
GX 1030