MI 3311 GammaGT
Autotest sequences
46
Reading appliance ID number with barcode reader or RFID reader/writer
When the instrument is in the Save results menu, appliance ID can be scanned from a
barcode label with the barcode reader or can be read from RFID tag with the RFID
reader/writer. A successful receive of the barcode or RFID tag is confirmed by two short
confirmation beeps.
6.5 Carrying out (Code, Simple and Custom) autotest
sequences
General meaning of keys during a Code, Simple and Custom autotest sequence:
TAB,
/
Sets option. Sets limit value in selected (highlighted) item.
ESC
Cancels autotest sequence and returns to the (Code, Simple and Custom)
autotest menu without changes.
TEST
Starts / repeats selected measurement or proceeds to next step.
Notes
If any of the inspections are marked as failed or if any test fails the test sequence is
stopped and the instrument automatically goes to the Result menu.
If a test parameter (limit, duration, output voltage) is changed the setup is valid only
for the particular test.
If the test limit, output voltage or test duration is changed in Code autotest
sequence the autotest code will not be stored (as the settings do not correspond
with the code anymore).
6.5.1 Visual inspection
Measurement is described in chapter
5.2.1 Visual inspection
.
Figure 6.18: Visual test menu
Options in Visual test:
PASS / FAIL
To be applied manually.
6.5.2 Earth continuity resistance measurement
The test is offered if it is applicable according to the autotest setting. The Earth continuity
starting screen is displayed first. Measurement and options in Earth continuity starting
screen are described in chapter
5.2.2 Earth continuity resistance
.
Figure 6.19: Earth continuity starting screen
After the measurement is carried out the Earth continuity result screen is displayed.